32
Z31900_00_01
Technical Data
Pneumatic
i
Pressure supply above 7 bar will be automatically repressed to avoid damage to the
SampleXpress.
4.3 Operating Conditions
Environment
For the appropriate temperature see also the Bruker site planning guides on the BASH
CD (Bruker Advanced Service Handbook):
Data Value Unit
Operating pressure 4-7 bar
Compressed air requirement, minimum > 100 l/min.
Oiling, maximum 0.005 mg/m
3
Table 4.3 Pneumatic Connection Values
Data Value Unit
Temperature range 5-30 °C
Relative humidity at 31 °C, maximum < 80 %
Decreasing linear till relative humidity < 50% at 40
°C, maximum
< 50 %
Table 4.4 Operating Environment
Manual Bruker Part Number
Site Planning for AVANCE Systems 300-700 MHz (UM) Z31276
Site Planning for AVANCE Systems 750 -950 MHz (UM) Z31686
Table 4.5 Bruker Site Planning Guides