Z31900_00_01
Maintenance
i
If the Information screen (Figure 10.3) does not appear, send the new file on the USB
memory stick back to the customer service. The file contains information about the
device.
10.7.2 Firmware Upgrade
i
Note: Make sure that the power supply will not be cut off during the entire procedure.
Otherwise you will have to send for a service technician.
Make sure that the USB memory stick has enough free space and is FAT32 formatted.
1. Stop all automatic operations.
2. Plug the USB memory stick in the USB port at the right side of the control panel.
A screen appears and the indicator lamp switches over to yellow.
Figure 10.4 Service File
The application software searches service files on the USB memory stick. This may
take several minutes.