Operation
106 Instructions for use Perseus A500 SW 2.0n
Customizing the current view
The monitoring area can be customized during
operation:
1 Touch a waveform or parameter field. The
Quick setup window opens.
2 For Content (A), select the desired content
from the list.
For a list of the possible screen content, see
page 173.
3 For Scale (B), select the desired scale.
Restoring the current view
The changes in the current view can be reset to
the saved standard.
1 Open the Views dialog window.
2 Touch the Restore current view button.
Using loops
The following loops are available:
– Pressure-Volume loop
– Flow-Volume loop
1 Open the Views dialog window.
2 Touch the Loops button.
The following is displayed:
A Current loop and the last 5 loops
B Parameters:
– Cdyn
– R
– TC
The area (E) can be configured so that the
Pressure-Volume loop is displayed.
10402
WARNING
Risk of insufficient monitoring
National and medical regulations require
certain parameters to be displayed.
Always consider the relevant regulations
when configuring the screen layout.
1047631735