Assembly and preparation
76 Instructions for use Perseus A500 SW 2.0n
Connecting or exchanging consumables
Single-use CO2 absorber
Connect or replace the CLIC absorber or the
Infinity ID CLIC absorber.
Strictly observe the instructions for use of the
single-use CO
2 absorber.
Reusable CO2 absorber
As an alternative to single-use CO2 absorbers, the
reusable CO
2 absorber may also be used.
Dismounting and emptying
1 Turn the CO
2 absorber (A) clockwise and
remove it from below.
2 Remove and dispose of the disposable dust
filter (B).
3 Empty used soda lime and dispose of
according to the instructions for use.
4 If it is necessary to clean the absorber insert
(C), remove the absorber insert from the
absorber container. Leave the inner and outer
sealing rings on the absorber insert.
WARNING
Risk of high inspiratory CO2 values
Use of soda lime over prolonged periods can
increase the inspiratory CO2 values.
Check the color of the soda lime regularly
and exchange as needed, especially if the
inspiratory CO
2 value increases
unexpectedly.
WARNING
Risk of insufficient ventilation
If the CO
2 absorber is not correctly locked
into place, system leakage may occur.
After mounting and replacing, make sure the
CO
2 absorber is firmly locked into place.
CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for
eyes, skin, and airway.
Handle this absorption material carefully and do
not spill it.
1034810349
A
B
C