Technical data
252 Instructions for use Perseus A500 SW 2.0n
Recommended separation distances
from wireless communication devices
To ensure that the full functional integrity of this
device is not compromised, there must be a
separation distance of at least 1.0 m (3.3 ft)
between this device and wireless high-frequency
communication equipment.
Emission of high-frequency energy
This medical device is equipped with an RFID
module (radio frequency identification) to enable
wireless communication with Infinity ID
accessories.
This medical device has been designed and
manufactured to comply with emission limit values
for high-frequency energy. These limit values are
incorporated in international safety standards such
as IEC 60601-1-2 and standards for radio
equipment such as EN 300330 and have been
defined by regulatory authorities.
The RFID system of this medical device complies
with Part 15 of the FCC regulations and the
license-free RSS regulations of Industry Canada.
Operation is subject to the following 2 conditions:
1 This medical device does not cause any
harmful interference.
2 The medical device is not liable to damage
caused by the reception of interference,
including interference causing undesired
operating conditions.
Changes and modifications that have not been
expressly approved by Dräger may result in the
user no longer being permitted to operate the
device.
Dräger hereby declares that this medical device,
including its radio equipment, is in compliance with
Directive 2014/53/EU.
The complete EU Declaration of Conformity can be
viewed at the following internet address:
http://www.draeger.com/doc-radio
Conducted high-frequency disturbances
(IEC 61000-4-6)
150 kHz to 80 MHz: 3 V, ISM bands: 6 V
Electromagnetic fields in the vicinity of wireless
communication devices
Various frequencies from 385 MHz to 5785 MHz:
9 V/m to 28 V/m
Immunity against Test level and required electromagnetic environ-
ment