Instructions for use Perseus A500 SW 2.0n 85
Operation
Safety information
WARNING
Risk of strangulation
Negligent placement of hoses, cables, sample
lines, and similar device components can
endanger the patient.
Use particular caution when establishing
connections to the patient.
CAUTION
Risk due to falsified gas measurement values
and failure of the gas measurement
Aerosols can damage the patient gas
measurement module and the membrane of the
water trap. Propellants can falsify the gas
measurement values.
Do not use medication nebulizers.
WARNING
Risk of faulty gas delivery
O
2 and CO2 and anesthetic gases must be
monitored every time breathing systems are
used.
Ensure adequate gas monitoring.
WARNING
Risk of misinterpretation
Misdiagnosis or misinterpretation of the
measured values or other parameters can
endanger the patient.
Do not make therapeutic decisions based
solely on individual measured values and
monitoring parameters. Therapeutic
decisions must be made solely by the user.
CAUTION
Risk of injury
Looking directly into the LEDs of the workplace
illumination and of the illumination unit on the
vaporizer plug-in connection can cause damage
to the retina.
Do not look directly into the LEDs. Make sure that
the patient is not dazzled by the LEDs.
CAUTION
Risk due to unsuitable illumination
If illumination without neutral colors is used
during the medical examination of the patient,
this may result in, e.g., misinterpretation of skin
coloring.
Do not use the device's workplace light during
examinations.
For examinations, use an examination light
conforming to IEC 60601-2-41.