Operation
116 Instructions for use Perseus A500 SW 2.0n
Ending the therapy
Switching to the Standby mode
1 In the main menu bar, touch the Standby...
button.
Or
Press the key below the screen.
2 In the Standby dialog window, confirm the
automatically selected Standby button.
With mechanically controlled gas mixer:
3 Close the flow control valves.
Using O2 insufflation
Electronically controlled gas mixer:
Mechanically controlled gas mixer:
WARNING
Risk of fire
In combination with oxygen or nitrous oxide,
ignition sources such as electrosurgical
devices and laser surgical devices can cause
fires.
Prevent leakage, e.g., at endotracheal
tubes, laryngeal mask airways, face
masks, Y-piece, breathing system
including hoses, filters, and breathing
bag, at the external fresh-gas outlet, and
at the outlet for O
2 insufflation.
Use only intact and leak-free hoses at the
outlet for O2 insufflation.
Before beginning laser surgery or
electrosurgery, flush with sufficient air
(<25 % O
2), and flush beneath the surgical
drapes as well.
Close the flow control valve on the O2
flowmeter to the end stop.
When O2 outlets are in use (e.g., for
insufflation), do not use any ignition
sources in the immediate vicinity.
Do not position oxygen sources in the
vicinity of ignition sources, e.g., electrical
connectors.
WARNING
Risk of a device fire
If there is a fire on the patient side, the device
may ignite.
If there is a fire, disconnect the device from
all oxygen-carrying supplies.
WARNING
Risk due to overpressure
When the patient is connected to the outlet
for O
2 insufflation without a relief valve,
increased pressure may be applied to the
patient.
Only connect the patient in a way that allows
excess gas to escape (e.g., through a relief
valve).
1035920364
B
C
A
B
A