Assembly and preparation
62 Instructions for use Perseus A500 SW 2.0n
When connecting D-Vapor 3000 vaporizers
1 Connect the power cable to an auxiliary power
socket beneath the column covering of
Perseus.
2 If needed, establish potential equalization.
3 If needed, stow the cable in the cable conduit.
Vapor View option
When combined with the Dräger Vapor 3000 or D-
Vapor 3000, the Vapor View option enables the
following functions:
Illumination of control dial and sight glass on
vaporizer
Reading the anesthetic agent type
Reading the control dial position
Monitoring the filling level
Prediction of the anesthetic gas concentration
Ensuring the gas scavenging
Perseus is equipped with an integrated anesthetic
gas receiving system (AGS).
1 Connect the scavenging hose to the nozzle (A)
on the receiving system.
2 Secure the scavenging hose with the strain
relief (B).
CAUTION
Risk of ambient air contamination
If the anesthetic gas receiving system is not
connected to the disposal system, contamination
of the ambient air with anesthetic gas may result.
Ensure the following:
The anesthetic gas receiving system is
correctly connected to the disposal system.
The anesthetic gas scavenging system is
functioning properly. Check the flow indicator.
10354
A
B