Instructions for use Perseus A500 SW 2.0n 41
Operating concept
Start values
Arrows on the scales of the therapy controls
mark the values resulting from the patient data and
start settings. The start values can be configured,
see page 150.
Linked therapy controls
Certain parameters can be linked to other
parameters. If one parameter is changed, the
linked parameter is also selected and changed.
Among other things, this applies to the adjustment
of ventilation pressures, ventilation times or during
electronically controlled fresh-gas delivery.
Example: The device can be configured so that a
change to the PEEP setting automatically causes
a change to Pinsp; as a result, the difference
between PEEP and Pinsp and therefore the tidal
volume remain constant.
Linking therapy controls, see page 162.
Additional information
Some settable parameters may be limited or be
mutually restricted so that certain combinations of
therapy settings are not possible, e.g., Ti 6.9 s at
RR 100 /min.
If a condition is reached in which a parameter
cannot be changed any more, Perseus displays a
corresponding message in the message field (D).
Dialog windows
Dialog windows consist of one or several pages
which are displayed by touching the corresponding
horizontal or vertical tab.
A Dialog window title
B Horizontal tab to open a page
C Vertical tab to open subordinate structures
D Button for closing the dialog window
Quick setup window
The Quick setup window is a context-sensitive
dialog window. Depending on the selected
parameter field or waveform, it contains various
setting possibilities, e.g., for limits, scale, or
content.
This window can be opened by touching the
corresponding parameter field in the monitoring
area. In the event of an alarm, the window can be
opened automatically, see page 155.
1040110402
A
B
C
D
Quick setup