Service
220 Instructions for use Perseus A500 SW 2.0n
It is recommended that only original parts from
Dräger are used and that the parts are replaced by
Dräger.
Repair
Repairs may be performed only by specialized
service personnel.
It is recommended that only original parts from
Dräger are used and that the parts are replaced by
Dräger.
Pressure reducer
(optional)
Cylinder pressure
reducer, type MP:
complete Every 6 years Replace Service personnel
Cylinder pressure
reducer, type MK:
Sealing Every 3 years
Sieve insert Every 3 years
Sintered metal filter
on the cylinder con-
nection
Every 3 years
Diaphragm unit Every 6 years Replace Specialized service
personnel
Valve cone Every 6 years
Blow-off valve Every 6 years
Safety valve spring Every 6 years
Component Interval Measure User group