Tests
130 Instructions for use Perseus A500 SW 2.0n
Breathing circuit
Prerequisite:
– Breathing system is complete and locked.
– Breathing system cover is in place.
1 Hoses (G) and filters, e.g., at the Y-piece (E),
are connected properly.
2 Extendable hoses (F) are extended to the
foreseen application length.
3 Remove the water from the hoses.
Valves
Checking the O2 flush
1 Press the O
2+ key and keep it pressed until the
pressure exceeds 15 hPa.
31900
NOTE
Do not change the length of flex hoses after the
test is done.
3189931901