Configuration
152 Instructions for use Perseus A500 SW 2.0n
System setup > Alarms
Vertical tab "Alarm limits"
Headline/
Parameter
Setting range
Description
Alarm limits for ventilation and gases for each patient category
FiO
2 [%]
19 to 99;
Off
19 to 99;
Off
19 to 99;
Off
90
Inspiratory oxygen fraction
FiO
2 [%]
18 to 98
20
18 to 98
20
18 to 98
20
etCO2
[%]; [kPa]
[mmHg]
0.1 to 9.8;
Off
7.0
1 to 75;
Off
53
0.1 to 9.8;
Off
7.0
1 to 75;
Off
53
0.1 to 9.8;
Off
7.0
1 to 75;
Off
53
Expiratory CO
2 concentration
etCO
2
[%]; [kPa]
[mmHg]
Off;
0.0 to 9.7;
Off;
0 to 74
Off;
0.0 to 9.7;
Off;
0 to 74
Off;
0.0 to 9.7;
Off;
0 to 74
inCO
2
[%]; [kPa]
[mmHg]
0.1 to 1.4;
Off
1.0
1 to 10;
Off
8
0.1 to 1.4;
Off
1.0
1 to 10;
Off
8
0.1 to 1.4;
Off
1.0
1 to 10;
Off
8
Inspiratory CO
2 concentration
Paw high
[mbar]; [hPa];
[cmH
2O]
5 to 99
40
5 to 99
25
5 to 99
20
Airway pressure
If the airway pressure is above the
set value for Paw low for longer
than 15 s, the alarm Airway press.
continuously high is issued. If the
setting Auto is selected, the lower
alarm limit is automatically
adapted to the set value for PEEP.
Paw low
[mbar]; [hPa];
[cmH
2O]
Auto;
3 to 97
Auto;
3 to 97
Auto;
3 to 97
inSev
[%]; [kPa]
0.10 to 9.95
4.40
0.10 to 9.95
5.10
0.10 to 9.95
6.70
Sevoflurane
inSev
[%]; [kPa]
Off;
0.00 to 9.85
Off;
0.00 to 9.85
Off;
0.00 to 9.85