Instructions for use Perseus A500 SW 2.0n 163
Configuration
System setup > Licenses/Options
Vertical tab "Licenses/Options"
Enable pause mode
(deactivates moni-
toring mode)
On; Off Select whether Pause mode or
Monitoring mode should be avail-
able.
On: Pause mode is available. The
mode has an adjustable timer,
after which an alarm is generated.
Pause can be activated from
standby and from all ventilation
modes.
Off: Monitoring mode is available.
Monitoring can be activated from
Standby mode and from Man/Spon
mode.
Headline/
Parameter
Setting range
Description
Licenses for software options Overview of available and
active software options.
Activating software options.
For more information, see
page 172
Gas supply
Disable N
2OOn; Off A device with connectors for
nitrous oxide can be configured so
that nitrous oxide is no longer dis-
played, nor can it be selected as a
carrier gas.
Prerequisite:
Current carrier gas is Air
Nitrous oxide is not connected
or available
To make the settings, restart Per-
seus if necessary.
On: Nitrous oxide cannot be deliv-
ered.
Off: Nitrous oxide can be deliv-
ered.