Instructions for use Perseus A500 SW 2.0n 141
Alarms
Suppressing the alarm tone
The alarm tone can be suppressed for a maximum
of 2 minutes.
Press the Audio paused key (A) below the
screen, see page 19.
The symbol and the remaining time for the
suppressed alarm tone are displayed in the header
bar.
After the alarm suppression has expired, the alarm
tone immediately resumes if the cause of the
alarm still exists.
While the alarm tone is suppressed, only those
alarms with a higher alarm priority or internal
priority rating are issued acoustically, see
page 191.
Reactivating the alarm tone
Press the Audio paused key again.
Downgrading and acknowledging alarm
messages
Some alarms can be downgraded to low priority or
can be cleared completely. The relevant alarms
can be identified in the table "Alarm – Cause –
Remedy" on page 191 by the following remedial
messages:
There are 2 options for downgrading or resetting
the alarms:
Option 1:
Or
Option 2:
After an alarm is downgraded, the respective
priority rating is maintained , see table "Alarm –
Cause – Remedy" on page 191.
10362
A
Remedial message Effect
Use "ALARM RESET"
to downgrade alarm
priority.
Alarm priority is
changed to low.
Use "ALARM RESET"
to acknowledge alarm.
Alarm is reset.
1049510488
Option 1 Option 2
Touch the ALARM
RESET button (A) in the
header bar and confirm.
In the Alarms > Cur-
rent alarms dialog
window (B), touch the
Reset all button (C) and
confirm.
All the alarms displayed
in the alarm message
field will be downgraded
or reset.
All alarms will be down-
graded or reset.
A
B
C
Alarms