Instructions for use Perseus A500 SW 2.0n 229
Technical data
Minimum pressure limit as per ISO 8835-5
and ISO 80601-2-13
0 hPa (cmH
2O)
Ventilator (continued)
Breathing system
Total volume including the disposable CO2
absorber CLIC Absorber Free
In the Man / Spon mode Typically 2.18 L
In automatic ventilation modes Typically 2.18 L
Compliance (without breathing circuit)
In the Man / Spon mode Typically 2.18 mL/hPa (mL/cmH
2O)
corresponds to 65.4 mL at 30 hPa (cmH
2O)
In automatic ventilation modes Typically 0.28 mL/hPa (mL/cmH
2O)
corresponds to 8.4 mL at 30 hPa (cmH
2O)
Soda lime filling volume for absorber container
Reusable CO
2 absorber, filled 1500 mL
Disposable CO
2 absorber CLIC absorber
800+
1300 mL
Disposable CO
2 absorber CLIC absorber
Free
1200 mL
Flexible arm for breathing bag
Volume 0.13 L
Compliance 0.13 mL/hPa (mL/cmH
2O)
corresponds to 3.9 mL at 30 hPa (cmH
2O)
Total system leakage <150 mL/min at 30 hPa (cmH
2O) standardized
to BTPS conditions
APL valve
Setting range Open, 5 to 70 hPa (cmH
2O)
Accuracy (in range from 15 to 30 L/min): ±20 % of the set value or ±3 hPa (the larger
value applies), but not more than +10 hPa
(cmH
2O)
Pressure drop at 30 L/min (ATPD), fully
opened
Dry: 2.1 hPa (cmH
2O)
Wet: 2.2 hPa (cmH
2O)