Service
216 Instructions for use Perseus A500 SW 2.0n
Transport for maintenance purposes
Additional information concerning the
ceiling-mounted version
For docking to the ceiling supply unit,
undocking from the ceiling supply unit, and for
transporting on the service cart, contact service
personnel.
Definition of service terminology
A service contract with Dräger is recommended.
WARNING
Risk of injury or damage to the device
If the medical device is not correctly docked
or has been undocked from the ceiling supply
unit, it may fall from the device mounting
interface or tip over.
The medical device may only be docked to
the ceiling supply unit or be undocked from it
by service personnel.
WARNING
Risk of injury or damage to the device
If the medical device was placed on the
service cart, it does not fulfill the tipping
stability requirements as per IEC 80601-2-13
and IEC 60601-1, and there is an increased
risk of tipping over.
– Only service personnel may move the
medical device on the service cart.
– Preparation for transport and actual
transport of the medical device on the
service cart must only be carried out if
there is no patient in the immediate
vicinity.
Concept Definition
Service All measures (inspection, mainte-
nance, repair) intended to maintain or
restore the functional integrity of a
product
Inspection Measures intended to determine and
assess the current state of a product
Mainte-
nance
Regular specified measures intended
to maintain the functional integrity of a
product
Repair Measures intended to restore the
functional integrity of a product after a
failure