Instructions for use Perseus A500 SW 2.0n 109
Operation
The entries in the logbook cannot be deleted and
are retained even after the device has been
switched off and on again or following a power
supply failure. When the storage limit is reached,
the oldest entries are overwritten.
Creating entries and associated settings, see
page 115.
Trends
Trends are displayed in the form of a graphic or a
table.
1 Open the Trends/Data dialog window.
2 Touch the Graphical trends (A) or Tabular
trends tab (B).
The following illustration shows the graphical
trend:
The vertical tabs (C) display different combinations
of parameters.
Zooming
In both trend displays, the displayed time period
can be enlarged or diminished.
Changing the time period:
Touch one of the following buttons (D)
according to choice:
Zoom +
Zoom -
Displaying the standard time period and the
current point in time:
Touch the Reset zoom button (E).
Moving the cursor
The exact measured values for a specific point in
time can be displayed numerically in area (F). To
do so, move the cursor.
Use the rotary knob to move the cursor (G).
Or
Touch the corresponding area on the screen.
Displaying installed options
Listing of the additionally installed software
options.
1 Open the System setup dialog window.
2 Touch the Licenses/Options tab.
Displaying an overview of the
accessories and consumptions
1 Open the System setup dialog window.
2 Touch the System status tab.
In Standby > System setup > System status >
Consumption, the gas consumption levels can be
reset to zero, see page 170.
10481
AB
C
D
E
F
G
Trends/Data
Vertical tab Overview
Accessories Accessories (when
Dräger Infinity ID accesso-
ries are used) and informa-
tion as to when the
accessory must be
replaced.
Supply Status display of the con-
nected gas and power sup-
plies.
Consumption Gas consumption
During operation:
for the current case
In Standby:
for the last case
since the last reset

Table of Contents