Instructions for use Perseus A500 SW 2.0n 189
Troubleshooting
Problems with the anesthetic gas receiving system (AGS)
Replacing the anesthetic gas receiving
system (AGS)
View from below:
Disassembling the anesthetic gas receiving
system (AGS)
1 Disconnect the scavenging hose.
2 If necessary, remove the transfer hose for the
non-rebreathing system.
3 Unfasten the 3 screws (A)
4 Dispose of the AGS.
Assembling the anesthetic gas receiving
system (AGS)
Follow the above steps in reverse order to
assemble the new AGS.
Fault Cause Remedy
Flow indicator beneath the
"restricted range"
The suction power of the ejector
in the terminal unit of anesthetic
gas scavenging system (AGSS)
is insufficient.
Have the function of the AGSS
terminal unit checked. Observe
related instructions for use.
Particle filter contaminated or
blocked.
Replace the anesthetic gas
receiving system (AGS) or have
service personnel replace the
particle filter.
Flow indicator above "normal
range"
The suction power of the ejector
in the AGSS terminal unit is too
high.
Have the suction power of the
ejector in the AGSS terminal unit
adjusted to the working range of
the AGS.
Particle filter missing. Replace the anesthetic gas
receiving system (AGS) or have
service personnel install the parti-
cle filter.