Assembly and preparation
78 Instructions for use Perseus A500 SW 2.0n
Connecting the sample line
Screw on the sample line at the Y-piece, HME
filter, or hose adapter and at the water trap.
Ensure the correct course of the sample line.
Do not use adapters.
WARNING
Risk of incorrect measured gas values
Blocked water traps or blocked sample lines
prevent correct gas measurement. As a
result, incorrect gas values may be displayed.
Use only Dräger sample lines.
Replace the O-rings of the water trap
mount every 2 years.
CAUTION
Risk of misleading data
Silicon can get into the measuring cuvette and
disrupt measurement.
Do not spray the O-rings of the water trap holder
with silicon spray.
WARNING
Risk of fire
Silicon or aerosol residues in the water trap
can cause fires.
Do not spray the O-rings of the water trap
holder with silicon spray.
WARNING
Danger due to incorrectly connected sample
line
If the sample line is connected to the
incorrect connections, e.g., infusion pump
connections, liquids instead of sample gas
can be sucked in. As a result, the gas
measurement will not display correct values.
When attaching the sample line, make sure
not to attach it incorrectly.
WARNING
Risk due to leakage from improperly
connected or damaged sample line
Connect the sample line correctly, otherwise
faulty gas measurements may result.
10344
WARNING
Risk of incorrect measured gas values
Blocked water traps or blocked sample lines
prevent correct gas measurement. As a
result, incorrect gas values may be displayed.
Use only Dräger sample lines.
Replace the O-rings of the water trap
mount every 2 years.