EasyManua.ls Logo

Dräger Perseus A500 - Screen fault;user interface failure; Complete failure

Dräger Perseus A500
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
188 Instructions for use Perseus A500 SW 2.0n
Screen fault/user interface failure
The screen does not respond to operation. It has
failed or the screen display is faulty.
1 Switch Perseus to Standby mode:
Press the button and confirm with the rotary
knob.
2 Use emergency O2 delivery, see page 183.
3 Check vaporizer setting.
4 Manually ventilate the patient.
5 Ensure corresponding substitute monitoring.
Complete failure
The device no longer responds to operation.
1 Turn off the device with the main switch, see
page 27.
2 Manually ventilate the patient.
3 Perform a start-up for operation in case of
emergency, see page 82.
WARNING
Risk of device malfunction
If the breathing bag does not fill with fresh
gas, the patient cannot be sufficiently
ventilated.
Check the oxygen supply and, if
necessary, open the gas cylinder valves.
If fresh gas still is not delivered or manual
ventilation is not possible, close the flow
control valve of the O
2 flowmeter.
Disconnect the patient from the device
and use a replacement device!

Table of Contents

Other manuals for Dräger Perseus A500

Related product manuals