Instructions for use Perseus A500 SW 2.0n 123
Tests
Performing the tests
Special characteristics of Dräger coaxial
hoses
When using coaxial breathing hoses, always use
the leakage assistant to check the inner hose of
the coaxial breathing circuit for leakage prior to the
system test or leakage test, see page 125.
System test and leakage test
Prerequisite:
Electronically controlled mixer: Central O
2
supply or central Air supply is connected.
Mechanically controlled mixer: Central O
2
supply is connected.
The anesthetic gas receiving system is
correctly connected.
The two tests consist of a checklist followed by an
automatic test.
The checklist can be presented in tabular form or
as a walk-through mode.
Button (D) can be used to switch from the tabular
checklist to the walk-through mode.
Whether tests always start in the walk-through
mode can be specified in the system setup, see
page 166.
1 Touch the button for the required test.
2 Complete the checklist (A), see page 126.
3 If all components are ready for operation, touch
the button (B). The automatic test will start.
If a component is not ready for operation, touch
the button (C).
The walk-through mode will start.
The components (E) are polled sequentially.
Buttons (F) are used to document whether the test
passed.
The automatic test starts after all the tests in the
walk-through mode are complete.
The test progress is displayed in area (G) while the
automatic test is running. All test results are
displayed in area (H).
10414
B
C
A
System test
System test - do not connect patient.
D
31887
Button Meaning
Test passed
Test failed
31888
System test - do not connect patient.
System test
E
FFFF
Test resultsSystem test
System test - do not connect patient.
G
H