Configuration
166 Instructions for use Perseus A500 SW 2.0n
Vertical tab "System test"
Headline/
Parameter
Setting range
Description
General
Always use walk-
through mode
On; Off When this function is switched on,
system tests and leakage tests will
always be executed in walk-
through mode.
Sample line is con-
nected during test
On; Off Specify whether the sample line is
connected to the Y-piece or to the
filter on the Y-piece during the
automatic tests.
When On is configured, the device
can automatically check whether,
e.g., breathing hoses or central
supply hoses are incorrectly con-
nected.
When Off is configured, additional
checks are required during the
system test.
Test gas supply
Central O
2 supply On; Off Select which gas supplies are
tested during the automatic sys-
tem test.
If Auto On is configured, only
check the gas supplies whose
actual availability is guaranteed
after the automatic start-up. Set-
ting both O
2 sources to Off - in
order to make the system test suc-
ceed - is not possible.
Central Air supply On; Off
Central N2O supply On; Off
O
2 cylinder On; Off
Air cylinder On; Off
N
2O cylinder On; Off
Verify O
2 delivery On; Off Select whether the system test will
check that the O
2 supply is actu-
ally delivering oxygen.
The Sample line is connected
during test parameter must also be
set to On for this.