Instructions for use Perseus A500 SW 2.0n 165
Configuration
Vertical tab "Units"
Vertical tab "Auto On/Flush & Dry"
Headline/
Parameter
Setting range
Description
Weight kg; lbs Set units.
Height cm; in
Airway pressure mbar; hPa; cmH
2O
Supply pressure bar; kPa×100; psi
CO
2 %; kPa; mmHg
Volatile agents %; kPa
Headline/
Parameter
Setting range
Description
Auto On Set the Auto On option. Perseus
provides a calendar function
which, when the device is shut
down, suggests the next time the
device should be ready for use,
depending on the weekday. To
activate the automatic ready-for-
use time for any desired day,
select the corresponding day and
set the time, see page 136.
Pay attention to the correct setting
of the gas supplies to be tested,
see page 166.
Day and time Monday to Sunday; Hour : Minute
Monday to Friday: 6:30
Saturday & Sunday: Off
When completed Standby/ Shut down
Flush & Dry
Flush duration
between cases
[min]
Off; 1 to 5
2
Set the flush cycle duration, e.g.,
at patient change.
Off:
Flush function is deactivated.
Flush & Dry dura-
tion before shut-
down [min]
Off; 5 to 30
10
Set the duration of flushing and
drying when the device is shut
down.
Off:
Drying function is deactivated.