Getting started
82 Instructions for use Perseus A500 SW 2.0n
Emergency start-up
Electronically controlled gas mixer
1 Adjust the APL valve.
2 Set O
2 switch (C) to the Add. O2 position.
3 Open the flow control valve (B) and set the
desired O
2 flow. If necessary, press the
O
2+ (D) key to quickly fill the breathing bag.
4 Monitor the set flow on the O
2 flowmeter (A).
5 Set the anesthetic gas concentration at the
vaporizer.
6 Manually ventilate the patient.
7 Switch on the device.
8 As soon as the Standby page is displayed,
start the therapy, see page 87.
9 Set the O
2 switch (C) to Aux. O2 to stop the
increased fresh-gas flow.
10 Close the flow control valve (B).
Mechanically controlled gas mixer
1 Adjust the APL valve.
2 Open the flow control valve (A) and set the
desired O
2 flow. If required, press the O2+ key
(C) to fill the breathing bag quickly.
3 Monitor the set flow on the total flow tube (B).
4 Set the anesthetic gas concentration at the
vaporizer.
5 Manually ventilate the patient.
6 Switch on the device.
7 As soon as the Standby page is displayed,
start the therapy, see page 87.
10376
D
C
B
A
20341
C
A
B