Instructions for use Perseus A500 SW 2.0n 231
Technical data
External fresh-gas outlet
Connection 22 mm outer taper / 15 mm inner taper (ISO)
Delivery See "Fresh-gas delivery"
Pressure limitation Not pressure-limited
Anesthetic gas receiving system
Suction flow
Normal range 32 to 50 L/min
At lower end of restricted range 14 L/min
Maximum fresh-gas flow to prevent contaminat-
ing ambient air
For external breathing systems (normal
range)
16 L/min
For external breathing systems (restricted
range)
5L/min
For internal breathing systems (restricted
range)
7L/min
Measuring systems and displays
As Perseus uses more accurate values for inter-
nal calculation and alarming, small deviations
due to rounding may occur.
Airway pressure
Airway pressure Paw
Plateau pressure Pplat
Positive end-expiratory pressure PEEP
Peak inspiratory pressure PIP
Mean airway pressure Pmean
Range –20 to +120 hPa (cmH
2O)
Accuracy ±4 % of the measured value or ±2 hPa (cmH
2O)
(the larger value applies)
Resolution of displayed value 1 hPa (cmH2O)