Instructions for use Perseus A500 SW 2.0n 237
Technical data
Accuracy ±4 % or ±6 kPa x 100 (the larger value applies)
±4 % or ±87 psi (the larger value applies)
±4 % or ±0.6 MPa (the larger value applies)
Resolution of displayed value 1 kPa x 100
1psi
0.1 MPa
Test results
Total leakage
Range 10 to 5000 mL/min
Resolution of displayed value 1 mL/min
Leakage in automatic ventilation
Range 10 to 5000 mL/min
Resolution of displayed value 1 mL/min
Hose compliance Compliance
Range 0 to 9.9 mL/hPa (mL/cmH
2O)
Resolution of displayed value 0.1 mL/hPa (mL/cmH
2O)
Measuring systems and displays (continued)
Display of calculated values
Measurement of consumption and production
CO
2 production of the patient
Range 0 to 9999 mL/min
Accuracy ±25 % or 100 mL/min (the larger value applies)
Resolution of displayed value 1 mL/min
O
2 uptake of the patient
Range 0 to 9999 mL/min
Accuracy ±25 % or 100 mL/min (the larger value applies)
Resolution of displayed value 1 mL/min
Fresh-gas consumption
Range 0 to 99999 L, only gas delivered by the gas
mixer is measured
Accuracy ±15 %
Resolution of displayed value 1 L