EasyManua.ls Logo

Dräger Perseus A500 - Pressure-Supported Ventilation

Dräger Perseus A500
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Principles of operation
262 Instructions for use Perseus A500 SW 2.0n
Pressure-supported ventilation
PSV (optional)
CPAP / PSV (optional)
Spontaneous breathing
Spontaneous breathing with continuous
positive pressure level with or without pressure
support
Each detected inspiratory effort at CPAP level
induces a patient-triggered, flow-controlled, and
pressure-supported breath. Point in time, number,
and duration of pressure-supported breaths are
determined by the patient. When no inspiratory
effort is detected, mechanical breaths are
delivered at the set minimum respiratory rate
RRmin and pressure support Psupp.
Patient-triggered breaths are ended as soon as the
inspiratory flow falls below the flow defined by the
Insp. term. setting. The duration of a machine-
triggered breath is additionally determined by the
patient category and the set minimum respiratory
rate RRmin.
Purely spontaneous breathing at CPAP level is
achieved by setting Psupp=Off. In this case the
patient does not receive any more machine-
triggered breaths.
10563
7LPH
7LPH
SDWLHQWWULJJHUHGEUHDWK
PDFKLQHWULJJHUHGEUHDWK
Paw
Slope
1/RRmin
Psupp
PEEP
Trigger
Insp. term.
Flow

Table of Contents

Other manuals for Dräger Perseus A500

Related product manuals