EasyManua.ls Logo

Dräger Perseus A500 - Response to Alarms; Alarm Priorities; Displaying Information on Alarms

Dräger Perseus A500
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alarms
140 Instructions for use Perseus A500 SW 2.0n
Alarm priorities
Perseus assigns the appropriate priority to each
alarm message.
The background color of the alarm message field
indicates the alarm priority of the active alarms.
The parameter field of the parameter triggering the
alarm flashes in the color matching the alarm
priority.
Response to alarms
Displaying information on alarms
1 Touch the alarm in the header bar.
Or
Open the Alarms dialog window and touch the
Current alarms tab (A).
2 In the list (B), touch the corresponding alarm or
select it with the rotary knob.
3 Refer to the information under Cause (C) and
Remedy (D) to remedy the error.
A list of all possible alarms can be found in chapter
"Troubleshooting", see page 191
Background
color
Alarm priority Marking Meaning
Red High !!! Immediate action is necessary in order to
avert imminent danger.
Yellow Medium !! Prompt action is necessary in order to
avert a danger.
Cyan Low ! Attention is necessary, but a delayed
response is sufficient.
10488
A
B
C
D
Alarms