Instructions for use Perseus A500 SW 2.0n 251
Technical data
EMC declaration
General information
This device was tested for electromagnetic
compatibility using accessories from the list of
accessories. Other accessories may only be used
if they do not compromise the electromagnetic
compatibility. The use of non-compliant
accessories may result in increased
electromagnetic emissions or decreased
electromagnetic immunity of the device.
This device may be used in the direct vicinity of
other devices only if Dräger has approved this
device arrangement. If no approval has been given
by Dräger, it must be ensured that this device
functions correctly in the desired arrangement
before use. The instructions for use for the other
devices must be followed.
Electromagnetic environment
This device may be used only in environments
specified in section "Environments of use" on
page 16.
Emissions Compliance
Radiated emis-
sions
Class A, group 1 (30 MHz to
1GHz)
Conducted emis-
sions
Class A, Group 1 (150 kHz to
30 MHz)
NOTE
The emissions characteristics of this equipment
make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11
class B is normally required), this equipment
might not offer adequate protection to radio-
frequency communication services.
The user might need to take mitigation
measures, such as relocating or re-orienting the
equipment.
Immunity against Test level and required electromagnetic environ-
ment
Electrostatic discharge (ESD) (IEC 61000-4-2) Contact discharge: ±8 kV
Air discharge: ±15 kV
Fast transient electrical disturbances (bursts)
(IEC 61000-4-4)
Power cable: ±2 kV
Longer signal input lines/output lines: ±1 kV
Impulse voltages (surges) (IEC 61000-4-5) Voltage, external conductor – external conductor:
±1 kV
Voltage, external conductor – protective ground
conductor: ±2 kV
Magnetic fields at mains frequency
(IEC 61000-4-8)
50 Hz: 30 A/m
Voltage dips and short interruptions in the sup-
ply voltage (IEC 61000-4-11)
Voltage dips of 30 % to 100 %, 8.3 ms to 5 s,
different phase angles
Radiated high-frequency disturbances
(IEC 61000-4-3)
80 MHz to 2.7 GHz: 3 V/m