Instructions for use Perseus A500 SW 2.0n 127
Tests
Vaporizers
For each vaporizer, check:
1 The locking lever points left, indicating the
vaporizer is locked.
2 Check the filling level in the sight glass. Refill
anesthetic agent if required.
When using a Dräger Vapor 3000 or
D-Vapor 3000 with the Vapor View option, a
yellow triangle on the screen indicates that the
filling level has fallen below the refill mark.
3 The control dial is set to position 0 and the key
is locked in placed.
4 The filling inlet is closed.