EasyManua.ls Logo

Dräger Perseus A500 - Accessories

Dräger Perseus A500
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Tests
132 Instructions for use Perseus A500 SW 2.0n
Manual resuscitator
1 Make sure there is a manual resuscitator on
the device.
2 Check the functional integrity of the manual
resuscitator.
Anesthetic gas receiving system
Have the flow for the anesthetic gas
scavenging system set so that the red flow
indicator (A) floats in range (B) ("normal
range").
When the flow indicator is floating in range (C)
("restricted range"), certain fresh-gas flow rates
should not be exceeded, see "Anesthetic gas
receiving system" in chapter "Technical data".
This can prevent contamination of the ambient
air.
Accessories
Soda lime
1 Make sure that the soda lime is not discolored.
Change the soda lime if it is discolored or when
its maximum period of use has been reached,
see page 76.
When Infinity ID functionality is configured:
Absorbers of the type Infinity ID CLIC
Absorber will automatically be detected,
and the exchange date will automatically be
set.
When Infinity ID functionality is not configured:
Absorbers will not be detected, e.g.,
reusable CO
2 absorbers
Update the exchange date manually: Touch
the Reset button after the soda lime has
been replaced.
3190631904
B
C
A
31922

Table of Contents

Other manuals for Dräger Perseus A500

Related product manuals