Operation
90 Instructions for use Perseus A500 SW 2.0n
Adjusting the therapy
Setting the APL valve
The pressure limitation set with the APL valve only
takes effect during manual ventilation or
spontaneous breathing.
The selection between manual ventilation
(Man) and spontaneous breathing (Spont) is
made at the APL pressure limitation valve, see
page 18.
Manual ventilation
Set the APL valve to the desired maximum
airway pressure.
The patient can be manually ventilated with the
breathing bag. The pressure is limited by the set
value.
In the Manual / Spontaneous mode, lifting the
valve relieves pressure from the breathing system.
Spontaneous breathing
Turn the APL valve counterclockwise as far as
it will go.
The points are vertically aligned and the valve is
raised.
The pressure limitation is canceled and the valve is
open for free spontaneous breathing.
WARNING
Risk of excessively high airway pressures
If the ventilator fails, the device switches into
the Man / Spon ventilation mode.
The APL valve should also be set to a
pressure limitation value suitable for the
patient when using automatic ventilation
modes. If the ventilator fails, ventilate the
patient manually.
10378
20 3010
1038010379

Table of Contents

Other manuals for Dräger Perseus A500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Perseus A500 and is the answer not in the manual?

Dräger Perseus A500 Specifications

General IconGeneral
BrandDräger
ModelPerseus A500
CategoryMedical Equipment
LanguageEnglish

Related product manuals