Instructions for use Perseus A500 SW 2.0n 219
Service
Maintenance
Component Interval Measure User group
CO
2 absorber/soda lime
with disposable dust fil-
ter
If colored violet or
according to the config-
ured Infinity ID exchange
interval
Replace, see page 76 Users
Water trap If required, if soiled, or
according to the config-
ured Infinity ID exchange
interval
Replace, see page 77 Users
Flow sensors If required, if calibration
is no longer possible, or
according to the config-
ured Infinity ID exchange
interval
Replace, see page 187 Users
AGS If required, if filter is
soiled, or flow is no lon-
ger achieved
Replace, see page 189 Users
AGS filter If required, if filter is
soiled, or flow is no lon-
ger achieved
Replace Service personnel
O-rings on the water trap
mount
Every 2 years Replace Service personnel
Sealing, sieve insert,
and sintered filter on the
cylinder connection of
the external cylinder
pressure reducer
(optional)
Every 3 years Replace Specialized service
personnel
Air filter mat
– Patient gas measure-
ment module
– Power supply unit
Every 3 years Replace Service personnel
CLIC adapter (optional) Every 4 years Replace Users
Lead-gel battery
(2 pieces)
Every 2 years Replace Service personnel