Configuration
164 Instructions for use Perseus A500 SW 2.0n
System setup > System
Vertical tab "General"
Headline/
Parameter
Setting range
Description
Language List of available languages
English (United States)
Select language.
A flag symbol identifies the tabs
that lead to the page with the lan-
guage settings.
Time source MEDIBUS 1; MEDIBUS 2; NTP server; None Select source for time synchroni-
zation.
Prerequisite: Connected device
supports this function.
IP address 4 numeric fields Enter the IP address of the NTP
server and confirm.
Date and time day; month; year
hour; minute
Set the date and time. The change
is applied on leaving the "General"
vertical tab.
Automatic switch to
daylight savings
time
On; Off Activate or deactivate automatic
switching to daylight saving time.
OR working hours Hour : Minute to Hour : Minute
6:30 to 18:30
Set the working hours of the oper-
ating room.
During this time, the gas measure-
ment is kept in a preheated and
calibrated state so measured val-
ues are available after only a short
waiting period. However, this
decreases the life span of the
patient gas measurement module.
Device name Rename
Device name (up to 16 alphanumeric charac-
ters)
Perseus
Change system name to, e.g.,
enter the installation site.
Configuration pass-
word
Change password Change the configuration pass-
word.
Reset all pages to Factory defaults Reset all the settings for the Sys-
tem setup dialog window to the
factory settings.