Instructions for use Perseus A500 SW 2.0n 143
Alarms
Adopting alarm settings when changing the ventilation mode
When the ventilation mode is changed, the alarm
settings are also adapted.
Depending on the mode, alarm settings can either
be adopted or set to Off.
Some modes can be configured whether or not the
settings are adopted.
However, the settings can be adjusted at any time
during operation.
Alarm or alarm
limit
Mode
VC,
VC - AF,
PC,
PC - APRV (optional)
PSV (optional) with
Psupp
5hPa
Pause, Man/Spon,
PSV (optional) with
Psupp <5 hPa
Ext. FGO
(optional)
CBM mode
FiO
2 low
Is restored or remains
active
Alarm settings are adopted
inAgent high
Apnea (no CO
2) Alarm settings are adopted
Off
etCO
2 high
etCO
2 low
configurable, see page 156
inCO
2 high
FiO
2 high configurable, see
page 156
inAgent low
xMAC low Not activated
Paw high
Alarm settings are
adopted
Not mea-
sured
Alarm settings are
adopted
Paw low
MV high
configurable, see
page 156
configurable, see
page 156
MV low
Apnea (no flow)
1)
1) This alarm is only activated when the MV low alarm limit is also activated.
Apnea (no pres-
sure)
On Off