Technical data
238 Instructions for use Perseus A500 SW 2.0n
Anesthetic agent
Range 0 to 999.9 mL liquid
Accuracy ±25 %
Resolution of displayed value 0.1 mL
Waveforms O
2 concentration
Primary anesthetic agent concentration
CO
2 concentration
Airway pressure
Volume (only for loops)
Flow
Sweep speed 6.25; 12.5; 25 mm/s
Scale
Airway pressure Paw (t) –20 to 80 hPa (cmH
2O)
Flow (t) –120 to 120 L/min
Volume (t) 0 to 2000 mL
O
2 (t) 0 to 100 Vol%
CO
2 (t) 0 to 100 mmHg (0 to 12 Vol%, 0 to 12 kPa)
Anesthetic agent (t)
Halothane 0 to 5 Vol% (kPa)
Enflurane 0 to 6 Vol% (kPa)
Isoflurane 0 to 5 Vol% (kPa)
Sevoflurane 0 to 10 Vol% (kPa)
Desflurane 0 to 20 Vol% (kPa)
Loops Pressure-Volume
Flow-Volume
Display of calculated values (continued)
Operating characteristics
Mains power supply
Electrical mains connection for the trolley
version
Without isolation transformer 100 to 240 V~ 50/60 Hz
With isolation transformer 100 to 127 V~ 50/60 Hz
Or
220 to 240 V~ 50/60 Hz