Instructions for use Perseus A500 SW 2.0n 239
Technical data
Maximum power consumption 12 A
Electrical mains connection for the ceiling-
mounted version
Ceiling-mounted version (without isola-
tion transformer)
100 to 240 V~ 50/60 Hz
Maximum power consumption 4 A
Power cable
Maximum length 5 m (16.4 ft)
Protective ground resistance Maximum 0.1 ohm
Operating voltage ≥250 V
Operating current ≥16 A
Power consumption
At 230 V~
Standby 0.30 A
Typical (without charging the internal
battery)
0.35 A
Maximum (with auxiliary power sock-
ets)
12 A
Maximum (ceiling-mounted version) 4 A
For 110 V~
Standby 0.55 A
Typical (without charging the internal
battery)
0.65 A
Maximum (with auxiliary power sock-
ets)
12 A
Maximum (ceiling-mounted version) 4 A
Power consumption
Standby 55 W
Typical 70 W
Maximum (trolley version) 2.2 kW
Maximum (ceiling-mounted version) 400 W
Inrush current (trolley version)
With isolation transformer Approx. 26 to 48 A peak
Approx. 18 to 34 A quasi-RMS
Without isolation transformer Approx. 8 to 14 A peak
Approx. 6 to 10 A quasi-RMS
Operating characteristics (continued)