EasyManua.ls Logo

Dräger Perseus A500 - Preparation for a day of operations; after reprocessing

Dräger Perseus A500
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use Perseus A500 SW 2.0n 63
Assembly and preparation
3 Connect the probe of the scavenging hose to
the terminal unit of the scavenging system.
As an option, the integrated anesthetic gas
receiving system can be operated in combination
with a control valve. Observe the assembly
instructions for the control valve.
Preparation for a day of operations / after reprocessing
Assembling the breathing system
1 Check all parts for damage or wear:
TurboVent 2 blower module
APL valve
Upper housing of the breathing system
Lower housing of the breathing system
Valve cages and valve plates
Flow sensors
–Ports
Incident flow meshes in the inspiratory limb
of the lower part of the breathing system
housing and in the expiratory port
Seals and sealing rings
2 The following parts must be free from deposits:
Incident flow meshes
Non-return valves (red and blue)
If necessary, remove deposits on the valve
plates of the non-return valves with a soft cloth.
If the incident flow meshes are damaged,
contact DrägerService.
Installing the TurboVent 2 blower module
1 Insert the TurboVent 2 blower module (A) into
the breathing system mount.
10353
WARNING
Risk of insufficient anesthetic gas
concentrations
If the component connections of the
breathing system are not leak-tight enough,
ambient air may be added to the anesthetic
gas mixture.
Make sure that all components of the
breathing system are connected tightly.
10540
A

Table of Contents

Other manuals for Dräger Perseus A500

Related product manuals