Instructions for use Perseus A500 SW 2.0n 217
Service
Inspection
Remote service
Perseus supports the following Remote Service
functions:
Help ticket
Remote Device Check
Proactive call
Contact the responsible DrägerService
organization for further information on the Remote
Service function.
Safety checks
Safety checks are no substitute for preventive
maintenance measures (including preventive
replacement of wear parts) as identified by the
manufacturer.
Performing the safety checks
1 Check whether the corresponding instructions
for use are available in the correct version.
2 Perform a functional test of the following
functions in accordance with the instructions
for use:
Emergency O
2 delivery
3 Check that the product is in good condition:
All labels are complete and legible
No visible damage to the following
components:
Trolley and castors
Housing parts
Brakes
Auxiliary power sockets
Vaporizer mount
Water trap mount
O-rings of the water trap mount
Screen
–Gas inlets
Status display
Breathing system
AGS and AGS valves
Hoses and cables
Strain reliefs for AGS, compressed gas
hoses, cables
Optional: Holding arms
Optional: Flexibility trolley, castors, and
locking mechanisms
Optional: Pressure reducers and their
sensor lines
Fuses which are accessible from the
outside are in compliance with the specified
values
Check that the country-specific labeling of
the gas type matches the screen display.
Check the end stop of the holding arms for
patient monitors.
4 Check the electrical safety in accordance with
the IEC 62353 standard.
5 Check the safety equipment:
Functional integrity of optical and acoustic
alarm generators
–O
2 switch on gas mixing unit (electronically
controlled gas mixer)
Internal battery
Optional: Check the function of the
minimum O2 delivery (mechanically
controlled gas mixer).
Measure Interval User group
Inspection and safety check Every 12 months Service personnel

Table of Contents