Configuration
156 Instructions for use Perseus A500 SW 2.0n
Vertical tab "Alarm config. 2"
Vertical tab "Alarm config. 3"
Headline/
Parameter
Setting range
Description
Deactivate the alarm limit when activating Man/Spon, CPAP,
CPAP/PSV with ∆Psupp < 5, or Pause?
The CPAP and CPAP / PSV
modes are optional.
FiO
2 high Yes; No Yes; No Yes; No Specifies the alarm behavior when
changing to a different ventilation
mode.
These settings apply only to a
change to a ventilation mode with
lower or no breathing support (see
page 259).
The alarm behavior at the start of
the therapy is defined by the con-
figuration in the vertical tab Alarm
limits.
MV low Yes; No
MV high Yes; No
xMAC low Yes; No
etCO
2 low Yes; No
etCO
2 high Yes; No
inCO
2 high Yes; No
inAgent low Yes; No
Deactivate the alarm limit in cardiac bypass mode (CBM)?
FiO
2 high Yes; No Specify alarm behavior in CBM
mode.
MV low Yes; No
MV high Yes; No
inAgent low Yes; No
Headline/
Parameter
Setting range
Description
Alarm limits for "Cylinder almost empty"
O
2 [bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
Specify alarm limits for supply
pressure of connected gas cylin-
ders.
Air [bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
N2O [bar]; [kPax100]: Off; 15 to 40
20
[psi]: Off; 218 to 580
290