Instructions for use Perseus A500 SW 2.0n 155
Configuration
Vertical tab "Alarm config. 1"
Headline/
Parameter
Setting range
Description
General alarm behavior
Open "Quick setup"
if alarm occurs
On; Off Opens the Quick setup window
automatically in the event of an
alarm.
"Second agent
detected" alarm
On; Off Issues an alarm when an anes-
thetic gas mixture is detected.
"xMAC low" alarm On; Off Activate the xMAC low alarm.
"FiO
2 too high for
neonates" thresh-
old value [%]
50; 25 to 90 Set the FiO
2 value that is consid-
ered critical for neonates. If this
value is exceeded for a certain
time, the FiO
2 too high for neo-
nates alarm will be triggered.
"FiO2 too high for
neonates" alarm
after [h:mm]
Off; 0:10 to 9:50
0:15
Set the time after which the FiO2
too high for neonates alarm will be
triggered.
Priority of "Apnea
ventilation" alarm
Medium; Low Specify the alarm priority when the
set minimum respiratory rate is not
reached in PSV ventilation mode.
Alarm behavior in "Pause" mode
Priority of "Pause
time expired" alarm
High; Medium; Low Specify the alarm priority for the
alarm that is issued when the time
set in the Pause mode has
expired.
Default value for
"Timer" [mm:ss]
0:30 to 02:00
2:00
0:30 to 02:00
1:00
0:30 to 02:00
0:30
Specify default time for Pause.