Instructions for use Perseus A500 SW 2.0n 167
Configuration
Vertical tab "Logbook"
Vertical tab "Sound volume"
Test breathing circuit
Test correct assem-
bly of breathing
hoses and Y-piece
On; Off Select whether the automatic test
will check that the breathing circuit
is correctly connected.
During this test, a check is made
as to whether the breathing gas
can flow from the inspiratory port
via the Y-piece to the expiratory
port.
If Sample line is connected during
test is set to On, this test is carried
out fully automatically. If Sample
line is connected during test is set
to Off, additional manual checks
are required.
Headline/
Parameter
Setting range
Description
A logbook entry with measured values is created Create additional logbook entries
with measured values.
Every 1 min; 2 min; 5 min; 10 min; 15 min Create entries regularly.
For all high-priority
alarms
On; Off Create entries for alarms.
For all medium-pri-
ority alarms
On; Off
Headline/
Parameter
Setting range
Description
Alarm tone volume 10 to 100
40
Set the alarm tone volume.
Minimum alarm tone
volume
10 to 100
10
Set the minimum volume with
which an alarm tone will be sig-
naled.
Breathing sound
volume (optional)
Off; 10 to 100
50
Set the breathing sound volume.
Headline/
Parameter
Setting range
Description