Configuration
168 Instructions for use Perseus A500 SW 2.0n
Vertical tab "Interfaces"
Headline/
Parameter
Setting range
Description
LAN Configure the network.
DHCP On; Off Make network settings.
When using DHCP, consult with IT
personnel to ensure that Perseus
is always assigned the same IP
address by the DHCP server.
Restart the device after each
change to the network settings.
The network settings are not
affected by a reset to factory set-
tings.
Accept the changes to IP address,
Subnet mask, or Default gateway
with the Apply button. The
changes are only active after the
device has been restarted.
IP address XXX . XXX . XXX . XXX
Subnet mask XXX . XXX . XXX . XXX
Default gateway XXX . XXX . XXX . XXX
MAC address Displays the MAC address.
COM 1 Configure the COM interfaces.
A baud rate of 19200 or 38400 is
required for transmission of high-
speed data, e.g., for waveforms.
Protocol MEDIBUS.X; None
Baud rate 1200; 2400; 4800; 9600; 19200; 38400
COM 2
Protocol MEDIBUS.X; None
Baud rate 1200; 2400; 4800; 9600; 19200; 38400
USB Activate or deactivate the USB
interface.
USB interface On; Off