Instructions for use Perseus A500 SW 2.0n 169
Configuration
Vertical tab "Infinity ID"
Vertical tab "Service"
Headline/
Parameter
Setting range
Description
Monitoring of Infinity ID accessories
Breathing circuit On; Off Activate or deactivate the
Infinity ID functionality.
On:
– Generates a message when
the maximum period of use is
exceeded
– Generates a message when
Infinity ID breathing hoses are
incorrectly connected
Off:
– Messages are suppressed.
Water trap On; Off
Flow sensors On; Off
CO2 absorber On; Off
Exchange interval [days]
Breathing circuit Off; 2 to 9
2
Set the exchange intervals for
Infinity ID accessories.
Water trap Off; 28
Flow sensors Off; 1 to 180
90
CO
2 absorber Off; 1 to 28
7
Headline/
Parameter
Setting range
Description
The following functions are avail-
able after the appropriate creden-
tials have been entered:
– Access to the service dialog