Instructions for use Perseus A500 SW 2.0n 153
Configuration
Vertical tab "Alarm tone volume"
inDes
[%]; [kPa]
0.1 to 20.0
12.5
0.1 to 20.0
14.5
0.1 to 20.0
19.0
Desflurane
inDes
[%]; [kPa]
Off;
0.0 to 19.9
Off;
0.0 to 19.9
Off;
0.0 to 19.9
inEnf
[%]; [kPa]
0.10 to 9.95
3.60
0.10 to 9.95
4.10
0.10 to 9.95
5.40
Enflurane
inEnf
[%]; [kPa]
Off;
0.00 to 9.85
Off;
0.00 to 9.85
Off;
0.00 to 9.85
inIso
[%]; [kPa]
0.10 to 8.50
2.40
0.10 to 8.50
2.80
0.10 to 8.50
3.70
Isoflurane
inIso
[%]; [kPa]
Off;
0.00 to 8.40
Off;
0.00 to 8.40
Off;
0.00 to 8.40
inHal
[%]; [kPa]
0.10 to 8.50
1.60
0.10 to 8.50
1.90
0.10 to 8.50
2.40
Halothane
inHal
[%]; [kPa]
Off;
0.00 to 8.40
Off;
0.00 to 8.40
Off;
0.00 to 8.40
Headline/
Parameter
Setting range
Description
Alarm tone volume 10 to 100
40
Set the alarm tone volume.
Minimum alarm tone
volume
10 to 100
10
Set the minimum volume with
which an alarm tone will be sig-
naled.
Headline/
Parameter
Setting range
Description
Alarm limits for ventilation and gases for each patient category