Instructions for use Perseus A500 SW 2.0n 233
Technical data
Low-flow wizard
Range 0 to 8 L/min
Accuracy ±25 % of measured value or ±100 mL/min (the
larger value applies)
Respiratory rate
Total RR
Spontaneous RRspon
Mandatory RRmand
Range 0 to 150 /min
Accuracy ±1 /min or 10 % (the larger value applies)
Resolution of displayed value 1 /min
T
0...90 <45 s (RR 6/min)
<105 s (RR <6 /min)
Dynamic compliance
Compliance Cdyn
Mean compliance Cdyn mean
Range 0 to 200 mL/hPa (mL/cmH
2O)
Accuracy ±30 % or ±3 mL/hPa (mL/cmH
2O) (the larger
value applies)
If during automatic ventilation there is still an
inspiratory flow at the end of inspiration, the
device will determine how much volume could
not be applied. This volume will then be
included with the normal volume in the compli-
ance calculation.
With increasing spontaneous breathing activity,
the compliance values can be distorted. In this
case the measurement accuracy may be
reduced.
Resolution of displayed value 0.1 mL/hPa (mL/cmH
2O)
Resistance R
Range 0 to 100 hPa / (L/s) (cmH
2O / (L/s))
Accuracy ±3 hPa / (L/s) (cmH
2O / (L/s)) or ±30 % of the
measured value (the larger value applies)
(With increasing spontaneous breathing activity,
the R values can be severely distorted. Hence
maintaining measurement accuracy cannot be
guaranteed with spontaneous breathing).
Resolution of displayed value 1 hPa / (L/s) (cmH2O / (L/s))
Measuring systems and displays (continued)