Instructions for use Perseus A500 SW 2.0n 65
Assembly and preparation
Mounting the upper part of the breathing
system housing
1 Place the upper part (G) of the breathing
system on the lower part (H).
2 Pay attention to the correct alignment of the
three levers (I), see illustration above.
3 Turn the 3 levers (I) by about 120° clockwise.
Inserting the flow sensors and the ports
1 Insert flow sensors (J) and (L).
2 Insert the inspiratory port (K) and screw it tight
with the knurled nut.
3 Insert in the expiratory port (M) and screw it
tight with the knurled nut.
10357
WARNING
Risk of fire
High temperatures arise in the flow sensors,
for example, in operation or while calibrating
during a system test. Due to the high
temperatures, residual vapors of flammable
disinfectants (e.g., alcohols) and residues
that were not removed during reprocessing
may ignite.
– Ensure particle-free cleaning and
disinfection.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor, check for
visible damage and soiling, such as
residual mucus, medication aerosols, and
particles.
– Replace flow sensors when damaged,
soiled, or not particle-free.
10358