2-6
Connecting Voltage Measurement Cables, continued
Measurement
cable with
optional fuse
voltage adapter
WARNING
To avoid the risk of electric shock or burns, always connect the safety (or earth)
ground before making any other connections.
WARNING
To reduce the risk of fire, electrical shock, or physical injury it is strongly
recommended to fuse the voltage measurement inputs.
Fuses must be located as close to circuit connection as possible to maximize
protection.
WARNING
For continued protection against risk of fire or shock hazard replace only with
same type and rating of fuse.
WARNING
Do not replace fuse again if failure is repeated. Repeated failure indicates a
defective condition that will not clear with replacement of the fuse. Refer
condition to a qualified technician.
Contact Dranetz Customer Service for more information on the fused voltage adapter.
Refer to Dranetz Information Sheet titled Fuse Voltage Adapter 1000V AC/DC FVA-
1K1 & FVA-1K4, P/N 899220.