Serinus® 10 User Manual 3.2
3. Once the Client application has loaded, click Home → Add Station → New Physical Station.
Figure 81 – Adding a New Station
4. This automatically brings the user to the Station tab on the ribbon. Enter the communication
details to connect to the instrument.
Figure 82 – New Station Connection
Table 8 – New Station Setup
The name of the station. If the user has other loggers, the name is used to distinguish
them.
Set this to “Ecotech Serinus” when downloading from any Serinus series instrument. This
will communicate with the instrument via the Advanced protocol. If using a network or
serial connection, ensure that the Advanced protocol has also been selected on the
instrument itself.
Set this to the time zone that the instrument is located in.
Enable this option if the user plan on changing the clock on the instrument with daylight
savings. Leave this disabled if the clock does not shift during DST. The instrument will
need to be adjusted manually for DST – it will not happen automatically.
This is the name to be used for the table in the SQL database containing this station’s
data. It must be unique for each station.