Serinus® 10 User Manual 3.2
Page 70
3.4.15 Diagnostics Menu
Figure 44 Diagnostics Menu Screen
Main Menu Service Menu Diagnostics Menu
Digital Pots Menu
Refer to Section 3.4.16.
Internal Pump Menu
[Internal Pump Option]
Refer to Section 3.4.17.
Valve Menu
Refer to Section 3.4.18.
Tests Menu
Refer to Section 3.4.19.
Pres/Temp/Flow Comp.
On (default): Is used to compensate instrument measurements for
environmental fluctuations that might affect readings (pressure,
temperature and flow).
Off: Is used only when running diagnostics.
Control Loop
Enabled (default): Allows the instrument to automatically adjust
digital pots and other outputs.
Disabled: Prevents the instrument from changing most outputs so
the service technician can manually control them.
3.4.16 Digital Pots Menu
Figure 45 Digital Pots Menu